Specifications |
Model 515C Oximeter
PRINCIPLE OF OPERATION
Red/Infrared absorption.
SpO2 (OXYGEN SATURATION)
| Range: |
0-100% |
| Accuracy: |
±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2 |
| Resolution: |
1% |
| Averaging: |
8 seconds |
PULSE RATE
| Range: |
30-250 bpm |
| Accuracy: |
±2 bpm |
| Resolution: |
1 bpm |
| Averaging: |
8 seconds |
SENSORS
Reusable Y-Sensor™ (can be sterilized and used with all patient
populations) and reusable adult finger sensor.
PLETHYSMOGRAM
Pulsatile waveform with autogain.
General
ALERT
Automatic and adjustable limits for SpO2
and Pulse Rate
| Audio: |
Adjustable volume, 2 min. silence or OFF (LED indicators) |
| Visual: |
Flashing numerics upon violated limit(s) & red "Alert Bar". Indication of sensor disconnect, sensor off patient, low signal, insufficient light, light interference, pulse out of range, sensor faulty and monitor faulty conditions. |
DISPLAY
| Numerics: |
7-Segment LED's |
| Pulse Bar: |
LED Pulse Bar Graph proportional to signal strength |
| Plethysmogram (Model 515C): |
Liquid Crystal Graphic module, 2.4" W × 1.3" H (6.0 x 3.3 cm) |
PHYSICAL
| Size: |
3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm) |
| Weight: |
6 lb (2.72 kg) |
ELECTRICAL
| Power requirements: |
100-120/200-240 VAC, 50-60 Hz, 30 VA |
| Battery: |
Sealed lead acid gel cell, 8 hr. life, 12 hr. recharge
Indicators for battery operation, low battery, and extremely low battery. |
ENVIRONMENTAL
| Operating temperature: |
50-104° F (10-40° C) |
| Operating humidity: |
0-90% relative (non-condensing) |
|