PRINCIPLE OF OPERATION
Red/Infrared absorption.

SpO2 (OXYGEN SATURATION)
Range: 0-100%
Accuracy: ±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
Resolution: 1%
Averaging: 8 seconds

PULSE RATE
Range: 30-250 bpm
Accuracy: ±2 bpm
Resolution: 1 bpm
Averaging: 8 seconds

SENSORS Reusable Y-Sensor™ (can be sterilized and used with all patient populations) and reusable adult finger sensor.

PLETHYSMOGRAM
Pulsatile waveform with autogain.


Top of Page

General

ALERT
Automatic and adjustable limits for SpO2 and Pulse Rate
Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
Visual: Flashing numerics upon violated limit(s) & red "Alert Bar". Indication of sensor disconnect, sensor off patient, low signal, insufficient light, light interference, pulse out of range, sensor faulty and monitor faulty conditions.

DISPLAY
Numerics: 7-Segment LED's
Pulse Bar: LED Pulse Bar Graph proportional to signal strength
Plethysmogram (Model 515C): Liquid Crystal Graphic module, 2.4" W × 1.3" H (6.0 x 3.3 cm)

PHYSICAL
Size: 3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm)
Weight: 6 lb (2.72 kg)

ELECTRICAL
Power requirements: 100-120/200-240 VAC, 50-60 Hz, 30 VA
Battery: Sealed lead acid gel cell, 8 hr. life, 12 hr. recharge
Indicators for battery operation, low battery, and extremely low battery.

ENVIRONMENTAL
Operating temperature: 50-104° F (10-40° C)
Operating humidity: 0-90% relative (non-condensing)


Top of Page

 


Respironics Home  |  Glossary  |   Careers  |  Online Stores  |  Warranties  |  Providers
Suppliers  |  Clinicians  |  Media Resource Library  |  Contact Us
All Rights Reserved  |  Privacy Notice  |  Terms & Conditions  |  Site Map

 

HOME

Features

Specifications

Part Numbers &
Accessories


Manuals

eLearning